Dilpreet Singh, Bhavuk Garg, Pulak Mohan Pandey and Dinesh Kalyanasundaram
The purpose of this paper is to establish a methodology for the design and development of patient-specific elbow implant with an elastic modulus close to that of the human bone…
Abstract
Purpose
The purpose of this paper is to establish a methodology for the design and development of patient-specific elbow implant with an elastic modulus close to that of the human bone. One of the most preferred implant material is titanium alloy which is about 8 to 9 times higher in strength than that of the human bone and is the closest than other metallic biomedical materials.
Design/methodology/approach
The methodology begins with the design of the implant from patient-specific computed tomography information and incorporates the manufacturing of the implant via a novel rapid prototyping assisted microwave sintering process.
Findings
The elastic modulus and the flexural strength of the implant were observed to be comparable to that of human elbow bones. The fatigue test depicts that the implant survives the one million cycles under physiological loading conditions. Other mechanical properties such as impact energy absorption, hardness and life cycle tests were also evaluated. The implant surface promotes human cell growth and adhesion and does not cause any adverse or undesired effects i.e. no cytotoxicity.
Practical implications
Stress shielding, and therefore, aseptic loosening of the implant shall be avoided. In the event of any trauma post-implantation, the implant would not hurt the patient.
Originality/value
The present study describes a methodology for the first time to be able to obtain the strength required for the medical implant without sacrificing the fatigue life requirement.