To read this content please select one of the options below:

Design and development of 3D printing assisted microwave sintering of elbow implant with biomechanical properties similar tohuman elbow

Dilpreet Singh (Centre for Biomedical Engineering, Indian Institute of Technology Delhi, New Delhi, India and Council of Scientific and Industrial Research, Central Mechanical Engineering Research Institute, Durgapur, India)
Bhavuk Garg (Department of Orthopedics, All India Institute of Medical Sciences, New Delhi, India)
Pulak Mohan Pandey (Department of Mechanical Engineering, Indian Institute of Technology Delhi, New Delhi, India)
Dinesh Kalyanasundaram (Centre for Biomedical Engineering, Indian Institute of Technology Delhi, New Delhi, India, and Department of Biomedical Engineering, All India Institute of Medical Sciences, New Delhi, New Delhi, India)

Rapid Prototyping Journal

ISSN: 1355-2546

Article publication date: 25 October 2021

Issue publication date: 26 January 2022

312

Abstract

Purpose

The purpose of this paper is to establish a methodology for the design and development of patient-specific elbow implant with an elastic modulus close to that of the human bone. One of the most preferred implant material is titanium alloy which is about 8 to 9 times higher in strength than that of the human bone and is the closest than other metallic biomedical materials.

Design/methodology/approach

The methodology begins with the design of the implant from patient-specific computed tomography information and incorporates the manufacturing of the implant via a novel rapid prototyping assisted microwave sintering process.

Findings

The elastic modulus and the flexural strength of the implant were observed to be comparable to that of human elbow bones. The fatigue test depicts that the implant survives the one million cycles under physiological loading conditions. Other mechanical properties such as impact energy absorption, hardness and life cycle tests were also evaluated. The implant surface promotes human cell growth and adhesion and does not cause any adverse or undesired effects i.e. no cytotoxicity.

Practical implications

Stress shielding, and therefore, aseptic loosening of the implant shall be avoided. In the event of any trauma post-implantation, the implant would not hurt the patient.

Originality/value

The present study describes a methodology for the first time to be able to obtain the strength required for the medical implant without sacrificing the fatigue life requirement.

Keywords

Acknowledgements

The work is funded by Indian Council of Medical Research (ICMR) under project no. I-1086 with file number 5/3/8/15/2019-ITR, Designing of patient specific total elbow replacement prosthesis for Indian Population.

Citation

Singh, D., Garg, B., Pandey, P.M. and Kalyanasundaram, D. (2022), "Design and development of 3D printing assisted microwave sintering of elbow implant with biomechanical properties similar tohuman elbow", Rapid Prototyping Journal, Vol. 28 No. 2, pp. 390-403. https://doi.org/10.1108/RPJ-05-2021-0116

Publisher

:

Emerald Publishing Limited

Copyright © 2021, Emerald Publishing Limited

Related articles