Sérgio B Gonçalves, Pedro Dantas, Francisco Guedes de Melo, João Gouveia, José Guimarães Consciência, Jorge Martins and Miguel Tavares da Silva

Arthroscopic osteochondroplasty is a minimally invasive procedure that has been used to treat femoroacetabular impingement syndrome, leading to significant improvements in…

Abstract

Purpose

Arthroscopic osteochondroplasty is a minimally invasive procedure that has been used to treat femoroacetabular impingement syndrome, leading to significant improvements in patients’ clinical outcomes and quality of life. However, some studies suggest that inadequate bone resection can substantially alter hip biomechanics. These modifications may generate different contact profiles and higher contact forces, increasing the risk of developing premature joint degeneration. To improve control over bone resection and biomechanical outcomes during arthroscopic osteochondroplasty surgery, this study aims to present a novel system for measuring femoroacetabular contact forces.

Design/methodology/approach

Following a structured design process for the development of medical devices, the steps required for its production using additive manufacturing with material extrusion and easily accessible sensors are described. The system comprises two main devices, one for measuring femoroacetabular contact forces and the other for quantifying the force applied by the assistant surgeon during lower limb manipulation. The hip device was designed for use within an arthroscopic environment, eliminating the need for additional portals.

Findings

To evaluate its performance, the system was first tested in a laboratory setup and later under in-service conditions. The 3D printing parameters were tuned to ensure the watertighness of the device and sustain the intraoperative fluid pressures. The final prototype allowed for the controlled measurement of the hip contact forces in real-time.

Originality/value

Using additive manufacturing and readily available sensors, the present work presents the first device to quantify joint contact forces during arthroscopic surgeries, serving as an additional tool to support the surgeon’s decision-making process regarding bone resection.