Theoretical and experimental analysis of a piezoelectric tactile sensor for use in endoscopic surgery
Abstract
This paper describes design, theoretical, and experimental analysis of a polyvinylidene fluoride (PVDF) tactile sensor, which could be integrated with an endoscopic grasper. The sensor exhibited high force sensitivity and linearity. Finite element analysis was employed to study the structural analysis of the tactile sensor with various load application and the results of this modelling are presented as the shear stress distribution and deformation contours. A comparison was made between the theoretical modeling and the experimental results.
Keywords
Citation
Dargahi, J. and Najarian, S. (2004), "Theoretical and experimental analysis of a piezoelectric tactile sensor for use in endoscopic surgery", Sensor Review, Vol. 24 No. 1, pp. 74-83. https://doi.org/10.1108/02602280410515851
Publisher
:Emerald Group Publishing Limited
Copyright © 2004, Emerald Group Publishing Limited